Contrast Media: 50 mL – 1000 mL Saline: 50 mL – 1000 mL
Reservoirs (volume)
3 x 200 mL reservoirs (two same API Contrast Media and one saline)
Air Detection
3 inlet air sensors to detect when bottles and bags run empty 1 outlet air sensor to detect air before it reaches the Patient Line Reservoir air detection and automatic air removal
Battery Performance
2 independent, Wi-Fi enabled, full color, touch screen units with modern graphical user interfaces located in the scan room and the control room
Barcode Reader
Integrated barcode reader on injector unit for contrast tracking and recording
Integrated Prime Contrainer (volume)
680 ml
Protocoling Specifications
Description
Specification
Volume Range
1 mL to 200 mL for contrast and saline phases; 1 mL to 400 mL for DualFlow phases (in 1 mL increments)
Flow Rate Range
0.1 to 10 mL/sec in 0.1 mL/sec increments
DualFlow
Simultaneous injection of contrast and saline in increments of 5%
Programmable Pressure Limit
From 50 psi to 300 psi in 1 psi increments (also displayable in kPa and kg/cm²)
Programmable Delay
1 second to 900 seconds in 1 second increments
Pause
Maximum pause time of 20 minutes
Injection Capabilities
Maximum of 6 phases per injection Maximum of 10 injections per exam
Skip to Next Phase
Enables the user to stop the current phase of an injection and start the next phase
Patency Check / Test Inject
Real time flow rate adjustment and multi-phase test injections
Disposables
Description
Specification
Patient Line
Single use, 250 cm, snap-in single hand design, 2 check valves, auto-primes upon insertion, recessed connection to prevent accidental touch
Day Set
Pre-assembled, labeled for 24-hour use, designed for set-up in less than 2 minutes, recessed connection to prevent accidental touch
Replacement Spikes
Labeled for 24-hour use, can be changed in case of an accidental touch, without having to change the entire Day Set
Workflow Solutions
Description
Specification
Automated Documentation
For access to contrast and injection information in PACS: Integrated barcode reader, Modality Worklist, PACS Interface
P3T algorithms calculate individualized injection protocols for cardiac CT, CT pulmonary angiography and CT imaging of abdominal organs Protocol Assistant provides basic protocol calculations for a range of procedure types
eGFR Calculator
Available on injector and Control Room Unit as part of the standard configuration
MEDRAD® Stellant
Learn more about the technical specifications of the MEDRAD® Stellant:
Syringe A: 0.5 mL to max syringe volume in: 0.1 mL increments up to 31 mL 1 mL increments above 31 mL Syringe B: 1 mL to max syringe volume in 1 mL increments
Programmable Pressure Limit (psi/kPa)
65 mL and 115 mL syringe pressure limit selections (PSI / kPa): Choice of 100/690, 150/1035, 200/1380, 250/1725, 300/2070, 325/2240 (default)
Programmable Flow Range (mL/sec)
0.01 to 10 mL/s in: 0.01 mL/s increments between 0.01 and 3.1 mL/s 0.1 mL/s increments between 3.1 and 10 mL/s
Pause
1-1200 seconds in 1 second increments
Hold
Maximum HOLD time is 20 minutes
Syringes (Volume capacity)
65 mL and 115 mL sterile disposable syringe
Electrical Requirements
100-240 VAC, 50-60 Hz, 1.3A
MEDRAD® Intego
Learn more about the technical specifications of the MEDRAD® Intego: